Coulometry is one of the rapid, most accurate, and precise analytical techniques that determine the quantity of an analyte by measuring the electrical charge needed for its complete electrolysis without using any analytical standards. The total charge passed during electrolysis correlates with the analyte amount by Faraday's laws of electrolysis. For accurate coulometric measurements, a charge equal to Faraday's constant multiplied by the number of electrons involved in the relevant half-reaction must be supplied per mole of analyte transformed at the electrode. This is represented by the equation:
Total Charge (Q) = n x F
where Q is the total charge required, n is the number of electrons involved, and F (Faraday's constant) is approximately 96,485 C/mol. A current efficiency of 100% is necessary for coulometric measurements.
Coulometric methods use either constant current or controlled potential. Constant-current coulometry measures the charge passed through the cell over time while applying a constant current. In contrast, controlled-potential coulometry applies a constant potential, measuring the charge passed through the cell as the current varies over time.
From Chapter 10:
Now Playing
Electrochemical Analyses and Redox Titration
556 Views
Electrochemical Analyses and Redox Titration
1.0K Views
Electrochemical Analyses and Redox Titration
728 Views
Electrochemical Analyses and Redox Titration
215 Views
Electrochemical Analyses and Redox Titration
952 Views
Electrochemical Analyses and Redox Titration
474 Views
Electrochemical Analyses and Redox Titration
398 Views
Electrochemical Analyses and Redox Titration
1.6K Views
Electrochemical Analyses and Redox Titration
1.1K Views
Electrochemical Analyses and Redox Titration
221 Views
Electrochemical Analyses and Redox Titration
633 Views
Electrochemical Analyses and Redox Titration
234 Views
Electrochemical Analyses and Redox Titration
181 Views
Electrochemical Analyses and Redox Titration
123 Views
Electrochemical Analyses and Redox Titration
148 Views
See More
Copyright © 2025 MyJoVE Corporation. All rights reserved