24.5K Views
•
16:40 min
•
July 31st, 2010
DOI :
July 31st, 2010
•The general goal of the following experiment is to determine the overall shape of protein complexes. This is achieved by acquiring mass spectrometry iron mobility data for protein complex ions and measuring the drift time values of each charge state. As a second step, the experimental conditions are validated in order to ensure mobility measurements of native protein structures.
Next, the measured drift time values are correlated with cross-sectional areas. Results determine the collision cross-section values of proteins or protein complexes with unknown three dimensional structures. This information provides clues on their overall shape, subunit packing and topology.
Hi, I'm Isaac Leski from the Lab of Mic, department of Biological Chemistry at the Wiseman Institute of Sciences, And I'm Naam Kirschenbaum also from the lab of mi. Today we'll show a procedure of measurement collision cross-sectional protein complexes using hybrid mass spectrometry and iron mobility instruments. We use this procedure in our laboratory to study the overall shape and ary organization of protein complexes.
So let's get started. This procedure focuses on iron mobility, mass spectrometry or IMM MS analysis of protein complexes. The sample preparation steps, instrument calibration and MS and tandem MS optimization procedures are demonstrated in a related JO protocol.
In general, this protocol involves low micromolar concentrations of complex in a volatile buffer such as ammonium acetate. Given that one to two microliters are consumed per nano flow capillary prepare 10 to 20 microliters as a minimum volume to enable optimization of MS conditions. To begin this procedure, set the traveling wave or T wave synap IMMS to the following modes of operation mobility tof in which the tri wave and pressures are automatically set for both IM and time of flight separation of ions, positive ion acquisition and V mode, setting the path of ions through the flight tube and reflection turn on all gases here.
Nitrogen is used for IM separation and argon for the trap and transfer regions. Recommended initial values are a gas flow of 24 milliliters per minute for the IMS device and 1.5 milliliters per minute for the trap region. Next, set the mast charge ratio acquisition range for an unknown protein complex.
Use a wide mass range initially, which can then be reduced to the desired values, adjust the MS profile accordingly. For maximum transmission efficiency for large complexes, the acquisition mass range should be set from 1030 2000 MA charge ratio and the MS profile to auto. Otherwise, the profile can be set according to the chart shown.
Check the RF setting and if necessary, adjust to values appropriate for large protein complexes as shown. Next, load the sample, apply capillary voltage and low nano flow pressure. Once sprays initiated, try to reduce the nano flow pressure to a minimal value.
In addition, adjust the position of the capillary with respect to the cone, adjust the MS acquisition parameters in order to acquire well resolved MS Spectrum. Optimize the pressure gradient along the instrument and the sampling cone, as well as the potential settings of the extraction cone bias trap and transfer as detailed in the associated JoVE protocol. Although these parameters are sample dependent shown, here are the conditions used for acquiring MS spectra of various iron masses from peptide to protein complexes.
Large ions require higher collision energies and bias voltage. It is also recommended to increase the back pressure for the analysis of large protein complexes to minimize the activation of the complex. Try to gradually reduce in steps of about 10 volts, the sample cone extraction, cone trap, and bias voltages without changing the position of the peak.
Once an optimum mass spectrum is obtained, the drift time or IM profile should be adjusted when analyzing protein assemblies. Optimal conditions for both mass and mobility measurements are often incompatible. Therefore, it is important to strike the proper balance between the two.
Overall, the iron mobility plot should be optimized such that the peaks are distributed over the entire drift time range and the peak profile is smooth. Approaching a gian distribution significant peak asymmetry can be related to poor separation of multiple conformations T-wave velocity, T-wave height, and IMS gas flow rate can be tuned to optimize the mobility separation. Increasing the T-wave velocity widens the drift time distribution profile.
While increased T-wave height values narrow it similarly increasing the IMS gas flow starting at 10 milliliters per minute minimum shifts the drift time profile toward higher values work to optimize the iron mobility spectrum. By fixing two of the three variables and optimizing the third, set the T-wave velocity to 250 meters per second and the gas flow to 24 milliliters per minute. Then as a starting point, set the height to three volts and in a stepwise manner, increase it in one volt increments.
When high bias voltages are used, it is recommended to reduce the IMS gas pressure and in this way enable the decrease in the bias voltage. As a consequence, complex activation and dissociation will be reduced. A rollover effect can appear when conditions are not optimized, observed as an identical peak in the first part of the drift times spectrum and the tailing edge.
When the ions do not traverse through the IAM device, effectively their journey may take longer than the time required for the next iron packet to be released into the mobility cell. As a result, a new bunch of ions is released from the trap region before the previous packet has been delivered to the pusher region. To eliminate this artifact, increase the T-wave height and decrease the T-wave velocity and IMS pressure.
In addition, the trap release time can be adjusted. Moreover, it is important to validate that the transfer T-wave height is set to at least five volts. Also to prevent leakage of ions towards the IMS cell.
The mobility trap height should be kept at maximum levels. Low velocity and high amplitude of transfer T waves may lead to the rippling of the drift time distribution profile. This artifact occurs when the mobility separation of ions is not maintained through the transfer and tough regions due to partial synchronization between the pusha frequency and the transfer T-wave velocity.
To eliminate this effect, either the pusher time or the transfer T-wave velocity should be adjusted. Since the pusha frequency is related to the mass range, this artifact may reappear. When this parameter is changed.
T-wave height exerts a minor effect. Though its reduction may also help to eliminate ripples. Once the aforementioned parameters are optimized, the IMMS data can be acquired to achieve highly resolved.
MS.Peaks protein complexes are often activated within the mass spectrometer to promote the stripping of residual water and buffer components. However, if the activation energy is increased beyond a threshold value, partial unfolding can occur forming multiple intermediate states, which are unlikely to correspond to the native solution state structure. As a result, the drift time peak may be shifted and broadened reflecting the hydrogenous population of unfolded structures.
In order to obtain drift time data consistent with solution phase structures, it is essential to carefully control the voltages used for accelerating ions prior to IM separation, therefore, increased capillary and cone voltage in a stepwise manner while monitoring the effect on the drift time spectrum. Moreover, for high MS resolution, it is preferable to increase the transfer rather than the trap voltage. The IM device is positioned first followed by the transfer region and the TOF analyzer.
Since the activation follows the IM measurement the remain unaffected for IM, while the MS accuracy can be increased to ensure that data acquisition is performed under conditions which maintain the native structure of the complex, it is important to collect data over a range of experimental and solution conditions rather than adhere to a single optimized set of parameters. For that reason, increase the trap collision voltage in a stepwise manner and acquire data at 10 volt intervals while monitoring the effect on the iron mobility profile. Finally, to identify the unfolded confirmations and assess the acquired data manually induce dissociation of the protein complex by titrating the sample with acetic acid over a pH range of two to seven and record the data proceed to analyze the data in the T wave IMS system, the cross-sectional areas defined by a drift time calibration approach, using Cain proteins with known cross-section values.
First, prepare denatured caliber protein solutions at 10 micromolar each. Use equine cytochrome C horse, heart myoglobin and bovine ubiquitin in 49, 49 0.2 volume ratio water methanol acetic acid. Next, acquire IMMS data for the caliber proteins under exactly the same instrument conditions used for the target protein or protein complex, keep all the voltages and pressure values identical to preserve the IM separation settings.
After acquiring the data, extract the experimental drift time value for each charge. State of the Cain proteins correct each of the caliber drift times T prime D using the following equation where MOVZ is the master charge ratio of the observed ion, and C is the enhanced duty cycle EDC delay coefficient. Its value typically between 1.4 and 1.6 is instrument dependent.
The EDC value is indicated within the system one acquisition settings, one acquisition setup tab, correct each of the caliber cross-sections for both iron charge state and reduced mass. Where omega C is the corrected cross-section, omega is the literature cross section Z is the iron charge. State M is the molecular weight of the cain ion, and MG is the molecular weight of the iron.
Background gas, which is typically nitrogen plop thelan of T prime D against thelan of omega C.The resulting curve corresponds to the following equation. The parameters X and A can be extracted by fitting the plot to a linear relationship. The slope X corresponds to the exponential proportion factor, and A represents the fit determined constant.
Calculate the fit correlation coefficient R squared. Acceptable values for R squared are greater than 0.95. A lower correlation coefficient value may be due to incomplete unfolding of the protein, aging of the sample.
Dissimilar experimental conditions used for the different caliber proteins, noisy spectrum and incomplete processing of data or calculation error. Correct the ca drift time using the determined exponential factor X and validate your calculations by relotting omega C versus T prime D.Define again, the correlation coefficient. Higher values than 0.95 are to be expected similarly to the steps described before, correct the measured drift time of the target protein or protein complex and calibrate the drift time of the target protein or protein complex using the calculated exponential factor X.Calculate omega of the target protein or protein complex using the fit determined constant A where omega equals a TD.Repeat these steps for each experimental condition.
When defining the cross-sectional area of the unknown protein or protein complex, we recommend that each experiment be repeated at least three times and the standard deviation of these triplicate measurements determined once the collision cross-section values are determined. Modeling approaches are employed in order to predict the topological or arrangements of the complex. This is done by fitting the experimental collision cross-section values within silico Omega values calculated from generated model structures on the whole, this field is still in its early years and further development is required to make this approach generic and applicable to a wide range of complexes.
Surface representation of the tetrameric form of bovine hemoglobin is shown here. The oxygen transporter hemoglobin complex can serve as an example for the approach mentioned above. Hemoglobin is a tetrameric protein complex composed of two alpha and two beta subunits, colored blue and red respectively, which form a dimer of alpha beta dimers.
The IMMS spectrum of hemoglobin is shown here. The acquired IM ms spectrum of the complex reveals a major charge series distribution corresponding to the intact complex and minor charge series, which fits the masses of the alpha beta dimer and the alpha and beta monomeric subunits. The theoretical and measured CCS of the different forms of hemoglobin is shown here.
Drift time values retrieved for several charge states of the di meric and tetrameric forms were used to calculate the omega values. These were compared to the theoretical values. Given that a tetrameric complex has three possibilities of association either cyclic dihedral or chain like packings.
By calculating the increase in omega when moving from a dimer to a tetramer, the structural organization can be predicted. Previous studies and our own experiments have shown strong correlation between measured ccss and protein surface areas derived from crystal structured data. This correlation can be used to calculate the expected increase in surface area of a tetramer compared to a dimer for the different packing forms.
This is done by considering each subunit aspheric object. A chain like assembly will increase the tetramer CCS at about twice while a C four or D two packing will yield an increase of about 1.5 and 1.67 respectively. The calculated ratio between the measured CCS values of the tetrameric and DME forms of hemoglobin was 1.57 plus or minus 0.03.
This number illustrates that the native structure is not linearly organized, but is arranged in a more compact form, either as a cyclic or dihedral tetramer. When repeating the same calculation on the resolved crystal structure of hemoglobin, the ratio of surface areas of tetrameric to DME forms was 1.63, which fits a D two symmetry. Obviously, this geometric model was simplified and protein subunits are not merely spherical.
However, this calculation demonstrates the exciting potential IMMS holds for revealing the packing of complexes with unknown high resolution structures. I've just shown you how to measure the drift time of the protein and protein complexes entail how to calculate their collision cross-section values. When doing this procedure, it is important to acquire the IMMS data for the caliber proteins using the exact same conditions used for the target protein or protein complexes.
Moreover, we strongly recommend repeating at least three time of these experiments and to determine standard deviation of these triplicate measurements. So that's it. Thank you for watching and good luck with your experiments.
イオンモビリティー質量分析法は、その衝突断面積と質量に基づいて、イオンを分離する新興気相技術です。方法は、タンパク質複合体の全体的なトポロジーと形状の三次元情報を提供します。ここで、我々は、機器設定や最適化、ドリフト時間のキャリブレーション、およびデータ解釈のための基本的な手順を概説する。
1:15
Acquiring Ion Mobility-Mass Spectrometry Spectrum
0:04
Title
0:46
Introduction
7:49
Screening Experimental Conditions
9:45
Correlating Between Drift Time Values & Cross-Sectional Areas
13:26
Representative Results
15:59
Conclusion
関連動画
19.3K Views
11.1K Views
12.5K Views
14.5K Views
16.0K Views
16.4K Views
17.4K Views
36.3K Views
15.8K Views
32.7K Views
Copyright © 2023 MyJoVE Corporation. All rights reserved